96P REAL-TIME PCR SYSTEM

Sensitive Photoelectric Detection System 1. High sensitivity photoelectric detector with no background noise which is fast and stable; 2. Lifetime maintenance-free ultra-bright LED cold light source w

Categories:

Sensitive Photoelectric Detection System

1. High sensitivity photoelectric detector with no background noise which is fast and stable;

2. Lifetime maintenance-free ultra-bright LED cold light source with a large signal value and high stability;

3. Unique optical fiber conduction technology that greatly improves fluorescence collection efficiency and avoids fluorescence interference between neighboring tubes;

4. The system automatically selects best gain value for each tube without the need for user configuration. The SLAN®has an extremely broad measuring range for fluorescence;

5. Extremely high detection sensitivity with an extremely low fluorescence background value. It’s not necessary for users to make optical correction or background correction.

 

SLAN-96P Multi-tasking Software Interface
(analysis screen)

 

Unique Electronic Automatic Hot Lid System

1. Flexible heating technology ensures uniform hot lid temperature;

2. The electronic automatic hot lid has an adjustment function for different 0.2ml reaction tubes to ensure even tube pressure;

3. Once the assay starts, the hot lid has automatic locking function to prevent assay failure following accidental opening.

 

Other functions

· Power-off protection · Automatic saving result · Export of result
· Real-time data display · Quantitative and qualitative analysis · File encryption
· Automatic baseline optimization · Digital filtering · Search function
· Report sheet print and edit · Melt curve function

Features

Precise Temperature Control System:

1. The SLAN®system adopts Peltier technology which can accurately and quickly control increases and decreases of temperature with a precision of ±0.1℃ and features long service life, no noise and no pollution, etc.;

2. Multiple-point temperature control technology ensures the uniformity between different wells. The temperature correction technology guarantees the accuracy of temperature and thus ensures the accuracy of assay result;

3. PCR tube temperature-control technology guarantees the accuracy of actual temperature of reagent in the tube and enables the SLAN®to easily adjust for different reactions using different volumes.

Specifications

Product model SLAN-96P
Sample capacity 96 wells (dual 48/48 reaction module)
Sample size 15-50mL
Consumable 0.2mL PCR tubes, 8-tube strips, 48-well plates
Parallel running mode Dual reaction blocks/running 2 tests independently
Temperature range 4-99°C
Max ramp rate 4.0°C/sec
Temperature accuracy ±0.1°C
Temperature uniformity ±0.1°C
Thermal cycling system Peltier-based system
Temperature control mode Tube control/block control
Light Source LED (maintenance-free)
Detector High sensitivity photoelectric sensor
Sensitivity 1 copy
Linearity range 100-1010
Resolution Can discriminate between 1000 copies and 2000 copies
Excitation CH1 470nm    CH2 530nm    CH3 580nm
CH4 630nm   CH5/CH6   custom-made
Emission CH1 510nm    CH2 565nm    CH3 620nm
CH4 665nm   CH5/CH6   custom-made
Dyes and probes CH1 FAMTM /SYBRGreen®   CH2 VIC®/HEX/J0ETM /TET
CH3 ROX/TexasRed®   CH4 CY5TM
Hot-lid 30°C~108°C (default105°C, adjustable) electronic automatic hot-lid
Power supply 230VAC, 50Hz
Power consumption 850VA
Dimensions 380mm x 520mm x 250mm (WxDxH)
Weight 18Kg
Communication RS232, USB
Qualification CE